We also tested the performance of DBS in the simulated 6 MV SL25
photon beam by examining fluence efficiency at the SSD (100 cm)
with the jaws widened to give a 40
40 cm
field.
The jaws simulated in this case were the ``standard" jaws
(10 cm thick, comprising 5 cm of tungsten and 5 cm of lead) instead
of the jaws for use with a multi-leaf collimator that were used to
determine performance of DBS with a 10
10 cm
field in this
accelerator (see section IV. above). Other than the jaws
and their settings, the simulation geometry and other parameters were identical
to those used with the 10
10 cm
field.
The splitting parameters used for all splitting routines were those
found to optimize performance in the 6 MV SL25 accelerator with a
10
10 cm
field (see section IV.B. above), with
the exception of the splitting field size, FS, in SBS and the splitting
field radius in DBS. For SBS, FS was set to 60 cm. This setting is
based on the performance of SBS as a function of FS in the
10
10 cm
beam (see section IV.B.2. above), where little
improvement in photon fluence efficiency was observed for
values of FS
field size + 20 cm. The splitting radius used
in DBS was 30 cm. This radius completely encloses the 40
40 cm
field, allowing for 2 cm beyond the corners of the field. Note that
with such a large splitting radius, the difference in efficiency with a
small change in the splitting radius (eg reducing it by 2 cm so that
it exactly encloses the field) is expected to be negligible.
For the purposes of scoring fluence, the phase-space surface at SSD=100 cm
was divided into 6561 (81
81) 1
1 cm
scoring zones. As
with the study of photon fluence efficiency in the 10
10 cm
beam, efficiency of all splitting algorithms is expected to increase
as the area of the scoring zones is decreased (see section IV.A.
above). Unlike the 10
10 cm
field case, the
1
1 cm
scoring
zones were used for both photon and electron fluence efficiency profiles.
|
It is clear from the figures that,
in the broad beam, DBS still offers a substantial improvement in efficiency
over the other splitting routines. In the case of photon fluence
within the field (
20 cm
X
20 cm), DBS is
between 5.5 (at the centre of the field) and 7 (at the edges of the
field) times more efficient than SBS and is
12 times more efficient
than UBS. Between
the edges of the field and the edge of the splitting field
(20 cm
X
30 cm), the
relative photon efficiency with DBS increases,
resulting in the ``horns" in Figure 11(a). This
increase is due to the high uncertainty (low efficiency) in the photon
fluence with no splitting in this region. In the case of electron fluence,
the efficiency using DBS is
8 times greater than with
SBS and
14 times greater than with UBS in the field.
The efficiency of DBS in the broad beam is significantly lower than
in the 10
10 cm
beam (Figure 6), with
photon fluence efficiency inside the field dropping by a factor
of
1.7 and electron fluence efficiency inside the field
dropping by a similar amount. In comparison, photon fluence efficiency
inside the field using SBS drops by a factor of only 1.1 (at the centre of
the field) to 1.5 (at the edges of the field) in the broad beam, with
electron fluence efficiency dropping by a factor of only
1.2.
In the case of UBS, the drop in photon and electron fluence efficiency
in the broad beam compared to the 10
10 cm
beam is insignificant.
The directional splitting routines (SBS and DBS) are less efficient in the broad beam simply because of the required increase in splitting field size. In the case of DBS, this results in both fewer photons being eliminated by Russian Roulette and more photons being generated by the do_smart_brems and do_smart_compton subroutines. In the case of SBS, this results in a higher splitting number over a greater range of incident electron directions/energies. The reason that the overall efficiency drop in the broad beam is relatively greater for DBS than for SBS may be due to the increased number of split Compton interactions in DBS (SBS does not split these interactions). In the case of UBS, the change in field size does not change the number of split photons that must be tracked, resulting in no significant efficiency change.
It is interesting to note that for a given splitting routine
in both broad beam and 10
10 cm
cases, the relative electron fluence efficiency is of the same order as
the relative photon fluence efficiency which is useful since electron
contamination plays a more important role in the broad beams.